To ensure accurate readings, calibrating your TENMARS TM-103 Sound Level Meter Data Logger using an Acoustic Calibrator is crucial. Here's a straightforward guide to help you calibrate your sound level meter effectively:
-
Prepare for Calibration:
- Ensure you have a standard Acoustic Calibrator that generates a 94dB, 1KHz Sine wave.
- Set your TENMARS TM-103 to dBA and Time weighting to FAST.
-
Insert the Microphone:
- Carefully place the microphone nozzle into the insertion hole of your Acoustic Calibrator.
-
Initiate Calibration:
- Press the A/C button on the sound level meter.
- Press and hold the HOLD button and the A/C button simultaneously for over a second. The LCD screen will begin blinking, indicating calibration mode.
-
Complete Calibration:
- Release the A/C and HOLD buttons while the LCD is blinking.
- If the calibration is successful, the sound level meter will display 94.0 dBA. If not, repeat the process until the meter shows the correct value.
- For optimal accuracy, recalibrate your TENMARS TM-103 annually.