Maintaining your TERRA UNIVERSAL Nitrogen Generator's performance requires timely filter element replacements. The frequency depends on your specific application and system configuration. Here's a breakdown:
- Prefilter: Inspect the pre-filter every three months. Replace it when the color indicator turns yellow. This helps protect the downstream components.
- Coalescing Filter: This filter also has a slide indicator. Monitor it monthly. Replace the coalescing filter when the indicator changes from green to yellow, and finally to red. Under typical continuous use at 100 PSI, these filters often last around two years or 20,000 hours.
Regularly checking these indicators and sticking to the suggested replacement schedule will keep your TERRA UNIVERSAL Nitrogen Generator running efficiently.