According to REOTEMP's guidelines, the frequency of calibrating your Back and Bottom Connect Thermometer is determined by your specific needs and application. To calibrate your REOTEMP thermometer, follow these steps: First, fully immerse the thermometer stem into a temperature-calibrated medium, ensuring at least 4 inches (100mm) of the stem is submerged. Next, use a small open-end wrench or a flat-tipped screwdriver to carefully turn the calibration adjustment screw located on the back of the thermometer case. This adjustment is designed for minor pointer corrections of less than 1%. Turn the screw to align the thermometer pointer with the correct temperature reading, if needed. Regular calibration is crucial for maintaining accuracy. Also, if your thermometer is installed without a thermowell, cleaning the stem periodically to remove any residue will help ensure accurate readings. For detailed information or further assistance, please contact REOTEMP directly or visit their website.