Setting the temperature on your IVYX Scientific -86°C Portable Deep Freezer is straightforward. Here's a step-by-step guide: First, press the SET key. The display will show "SP" (Set Point). Then, press the SET key again to view the current set temperature (SP value). You can adjust the temperature using the up and down arrow keys; the temperature can be set in 0.1°C increments, ranging from -40.0°C to -86.0°C. Press the SET key to confirm your desired temperature. The compressor will activate, and the freezer will begin cooling to the set temperature. Note that the red dot under "out" on the LCD screen indicates that the compressor is running. To exit the temperature setting mode, press the SET key and the up arrow key simultaneously. Alternatively, you can wait for one minute, and the current temperature inside the freezer will be displayed on the LCD screen. It's recommended to allow the freezer to reach the set temperature before adding samples. To minimize temperature fluctuations, avoid filling the freezer beyond ¼ of its capacity at once. Allow the temperature to stabilize for a few minutes after each addition of samples. For any further assistance or if you have any questions, please contact IVYX Scientific at support@ivyxscientific.com; they aim to respond within 24 hours.